A customer came to inquire that the DR exposure switch is not powered on, that is, the X-ray machine has no radiation when the exposure handbrake is pressed down. You can use a multimeter to detect whether there is voltage and current after the exposure handbrake button is pressed.
In this case, it may be that the wire of the exposure switch is broken, or the wiring of the exposure switch is faulty, and further testing is required. If the handbrake switch line is faulty, simply replace the new handbrake switch.
The hand-held switch is equipped with a two-stage grading switch. Press the button to turn on, and release it to turn off. The service life of the exposure switch is also very long, with an electrical service life of 300,000 times and a mechanical service life of 1,000,000 times.