Recently, Guangzhou ** Technology Co., Ltd. consulted the X-ray machine exposure hand switch of Huarui Image. They are used in medical rehabilitation equipment. They are currently in the research and development stage. Their parameters for the exposure hand brake are as follows: 1, a file 2 3. With self-locking 3. Withstand voltage greater than 220V and current greater than 10A. With reference to the customer’s parameters, our response is as follows:
1. The hand brake of a gear can use the L06 hand brake switch commonly used in dentistry (see the figure below) 2. The switch with self-locking needs special customization
3. The domestic standard for the voltage resistance and power supply of the hand brake is 36V-2A, and the parameter provided by the customer is greater than 220 volts and the current is greater than 10A. This parameter does not meet the safety factor of direct manual operation of the hand brake. It is recommended that the customer and the relevant departments implement the parameters of power supply and voltage.