Newheek x ray exposure hand switch 2-step What are the characteristics of [L01, L02, L04]?
• Good operating feel with a quick acting mechanism
• Easy to hold manually
• Long strokes through a unique mechanism
• Long-term experience with a variety of applications
●[L01, L02, L04] Wiring method
Newheek x ray exposure hand switch 2-step L01
3-core Step1 (Prepare Status): red and green wire
Step2 (Exposure Status): green and white wire
4-core Step1 (Prepare Status): black and white wire
Step2 (Exposure Status): red and green wire
Newheek x ray exposure hand switch 2-step L04
3-core Step1 (Prepare Status): red and green wire
Step2 (Exposure Status): green and white wire
4-core Step1 (Prepare Status): black and white wire
Step2 (Exposure Status): red and green wire


• X ray exposure hand switch is not sealed and cannot be cleaned. Doing so can cause the detergent to enter the switch with the flux or dust particles on the PCB, causing a malfunction.
●X ray exposure hand switch application/storage environment
• Do not use the switch in places where toxic gases, silicon, excessive dust, excessive dirt, high temperatures, high humidity, sudden temperature changes, water splashes or oil splashes. Otherwise, damage may occur due to contact contact failure, corrosion or other causes or other malfunctions.
• Check the operating characteristics, contact resistance, insulation resistance and dielectric strength of the switch before use if the following conditions exist;
• If stored for more than 1 year after production.
• If they fall down.
• If condensation occurs.
• If they are excluded from storage environmental conditions.